background: Commercially available ELISA-based antibody tests are used to approximate vaccination success against SARS-CoV-2 in at-risk patients, but it is unclear whether they correlate with neutralization of the Omicron variant.
methods: 269 serum samples of a cohort of 44 non-immunosuppressed participants and 65 MTX-treated rheumatic patients taken before and after COVID-19 booster vaccinations were measured using COVID-19 antibody testing systems with wild-type and Omicron BA.1 antigens developed by three different manufacturers (surrogate virus neutralization test cPass, and binding antibody tests QuantiVac and SeraSpot), as well as with a pseudovirus neutralization test (pVNT). The pVNT was considered the... More
background: Commercially available ELISA-based antibody tests are used to approximate vaccination success against SARS-CoV-2 in at-risk patients, but it is unclear whether they correlate with neutralization of the Omicron variant.
methods: 269 serum samples of a cohort of 44 non-immunosuppressed participants and 65 MTX-treated rheumatic patients taken before and after COVID-19 booster vaccinations were measured using COVID-19 antibody testing systems with wild-type and Omicron BA.1 antigens developed by three different manufacturers (surrogate virus neutralization test cPass, and binding antibody tests QuantiVac and SeraSpot), as well as with a pseudovirus neutralization test (pVNT). The pVNT was considered the gold standard for determining the presence and level of anti-SARS-CoV-2 antibodies.
results: All three wild-type ELISAs showed excellent test performance compared with wild-type neutralization in pVNT. However, out of 56 samples without Omicron BA.1 neutralization in pVNT, 71.4% showed positive results in at least one and 28.6% in all three wild-type ELISAs at the manufacturer-defined cut-offs. Omicron ELISAs showed either decreased specificity (57.1% and 55.4% for binding ELISAs) or sensitivity (51.2% in cPass) compared to Omicron neutralization in pVNT. The proportion of any false positive results among all samples decreased from 26.5% before to 3.2% after booster vaccination, however binding antibody test specificities remained below 70%.
conclusions: We found a poorer test performance of new Omicron antibody test systems compared to wild-type tests in detecting neutralizing antibodies against the corresponding SARS-CoV-2 variants. Decisions for booster vaccination or passive immunization of at-risk patients should not be based solely on antibody test results.